Cox, M G; Shirono, K (2023) Analysis of a regional metrology organization key comparison: model-based unilateral degrees of equivalence. Metrologia, 60 (5). 055014
Preview |
Text
eid9888.pdf - Published Version Available under License Creative Commons Attribution. Download (430kB) | Preview |
Abstract
A statistical testing method is developed to analyze a Regional Metrology Organization (RMO) key comparison (KC) with linking to the corresponding KC conducted by the International Committee of Weights and Measures (CIPM). We establish a statistical model to which a generalized least squares method is applied to estimate a linking invariant, ensuring that the CIPM KC reference value is unchanged by the analysis. The proposed method generally gives different unilateral and bilateral degrees of equivalence from those assessed by other methods. The approach accounts for uncertainty information provided by all participating laboratories and correlation information provided by the linking laboratories. Since decisions are made on the basis of calculated degrees of equivalence, we emphasize that it is valuable to have adequate knowledge of the properties of methods used for analyzing the results from an RMO KC.
| Item Type: | Article |
|---|---|
| Keywords: | CIPM and regional key comparisons, key comparison reference value, degrees of equivalence, statistical model, generalized least squares, uncertainty propagation |
| Subjects: | Mathematics and Scientific Computing > Measurement Uncertainties |
| Divisions: | Data Science |
| Identification number/DOI: | 10.1088/1681-7575/acf76c |
| Last Modified: | 05 Jan 2024 14:45 |