Bayliss, D J (2002) Measurement of the volume of a Zerodur sphere of nominal mass 1 kg by comparison hydrostatic weighing with Zerodur and silica standards of similar nominal mass. NPL Report. CMAM 80
|
Text
CMAM80.pdf Download (1MB) |
Abstract
This report describes a method for measuring the volume of a Zerodur sphere of nominal mass 1 kg by making comparison hydrostatic weighings with two calibrated volume reference standards of similar nominal mass manufactured from Zerodur and silica respectively. The result obtained was in good agreement with a values previously determined by the Physikalisch-Technische Bundesanstalt (PTB) and a UKAS accredited UK Laboratory.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | CMAM 80 |
| Subjects: | Engineering Measurements Engineering Measurements > Mass, Force, Pressure |
| Last Modified: | 02 Feb 2018 13:17 |