Gelat, P N; Hodnett, M; Zeqiri, B (2000) Establishing a reference ultrasonic cleaning vessel: Part 1 supporting infrastructure and early measurements. NPL Report. CMAM 55
|
Text
cmam55.pdf Download (602kB) |
Abstract
This document contains a description of the tasks completed within deliverable 4.5.1 "Establish a reference ultrasonic cleaning vessel" in the NMS Acoustics Programme 1998-2001. It includes the findings of a literature search, details of equipment specifications and performance, and the results of some initial hydrophone measurements carried out to assess the reproducibility of establishing a cavitating field using an ultrasonic cleaning bath.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | CMAM 55 |
| Subjects: | Acoustics Acoustics > Ultrasound |
| Last Modified: | 02 Feb 2018 13:17 |