Measurement report CEOS WGCV pilot comparison of techniques and instruments used for the vicarious calibration of land surface imaging through a ground reference standard test site 2009.

Behnert, I D; Deadman, A J; Fox, N P; Harris, P M; Gürol, S*; Özen, H*; Bachmann, M*; Boucher, Y*; Lachérade, S* (2011) Measurement report CEOS WGCV pilot comparison of techniques and instruments used for the vicarious calibration of land surface imaging through a ground reference standard test site 2009. NPL Report. OP 5

[thumbnail of OP5.pdf] Text
OP5.pdf

Download (6MB)

Abstract

This report presents the procedures and results of the CEOS pilot land comparison of techniques/instruments used for the vicarious calibration of optical remote sensors using as a reference standard test site Tuz Golu, Turkey. The measurement campaign participants were: DLR (German Remote Sensing Data Centre), ONERA (The French Aerospace Lab), CNES (French National Aerospace Agency), TU (Tubitak Space Technologies Research Institute, Turkey) and NPL (National Physical Laboratory, UK), where NPL was the pilot of this comparison. The report contains a detailed description of the instrumentation used by each participant, their calibration and traceability to SI standards, techniques used to characterise the site and the comparison of the results obtained by different teams. Each team estimated their uncertainty budget and the results are presented with the associated uncertainty separated into its components: type A and B.

Item Type: Report/Guide (NPL Report)
NPL Report No.: OP 5
Keywords: vicarious calibration, optical sensors, remote sensing
Subjects: International
International > Metrology
Last Modified: 02 Feb 2018 13:14
URI: https://eprintspublications.npl.co.uk/id/eprint/4933