Moreno, B; Million, M; Thomas, D; Taylor, G (2023) Assessment of the impact on the PADC neutron dosimetry system of the new operational dose quantities proposed in ICRU Report 95. Radiation Protection Dosimetry, 199 (15-16). pp. 1725-1728.
Full text not available from this repository.Abstract
The International Commission on Radiation Units and Measurements, ICRU, have recommended, in ICRU Report 95, the adoption of new operational quantities for use in radiation protection monitoring. Their implementation will pose problems for instrument manufacturers supplying area dosemeters, and for individual monitoring services providing personal dosemeters for radiation workers. The problems arise for photon, neutron, and beta dosimetry, but this paper deals solely with personal neutron dosemeters based on PADC plastic. It presents information on the problems to be expected based on data from calibrations of actual personal neutron dosemeters in terms of the existing dose quantities.
| Item Type: | Article |
|---|---|
| Subjects: | Ionising Radiation > Neutron Metrology |
| Divisions: | Medical, Marine & Nuclear |
| Identification number/DOI: | 10.1093/rpd/ncac212 |
| Last Modified: | 16 Feb 2024 14:51 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/9932 |
![]() |