Lewis, V E; McClure, D R* (1998) Second IRMF intercomparison of calibrations of portable gamma-ray dose-rate monitors, 1995-1996. NPL Report. CIRM 8
|
Text
CIRM8.pdf Download (13MB) |
Abstract
The Ionising Radiations Metrology Forum (IRMF) organised a second intercomparison of calibrations of gamma-ray dose-rate monitors in which sixteen establishments in the UK participated. The exercise involved the circulation of four gamma-ray monitors for calibration in the fields produced using 137CS, 6OCO and 241 Am. The instruments used were an NE technology PDM1B, a Mini-Instruments Mini-Rad (1000R), a Mini-InstrumE~nts Mini-900 with 44A Probe and a Siemens Electronic Personal Dosemeter.
All results were submitted to the authors for analysis. Responses relative to 'true' air-kerma rate, as stated by the individual participants, were calculated. The results, reported anonymously, are compared and demonstrate generally satisfactory agreement between establishments.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | CIRM 8 |
| Subjects: | Ionising Radiation Ionising Radiation > Dosimetry |
| Last Modified: | 10 Sep 2018 13:13 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/974 |
![]() |
Tools