Report on dose measurements on the MR-linac at Christie Hospital performed by NPL

Billas, Ilias; Duane, Simon (2018) Report on dose measurements on the MR-linac at Christie Hospital performed by NPL. NPL Report. IR 50

[thumbnail of NPL Report IR 50]
Preview
Text (NPL Report IR 50)
IR50.pdf - Published Version

Download (819kB) | Preview

Abstract

This report describes measurements of the radiation dose from the MR-linac at Christie Hospital using a selection of detectors (such as ionisation chambers and alanine) in order to calibrate the machine output. Absorbed dose to water calibration coefficients for the ionisation chambers used in the constant magnetic field (1.5 T) of the MR-linac were also determined. Together with similar measurements performed at NPL in a 0 T magnetic field over a range of beam qualities, the corrections required to measure the absorbed dose in the presence of a magnetic field using these detectors can be determined.

Item Type: Report/Guide (NPL Report)
NPL Report No.: IR 50
Subjects: Ionising Radiation > Dosimetry