Bass, G A; Duane, S; Galer, S E; Homer, M J; Kelly, M; Manning, J W; Maughan, D J; Nutbrown, R F; Sander, T; Shipley, D R; Snaith, J A D (2017) The NPL air kerma primary standard PS5-1/PS5-2 for 137Cs and 60CO: summary of factors. NPL Report. IR 41
Preview |
Text
IR41.pdf - Published Version Download (325kB) | Preview |
Abstract
Since the 1950s the UK primary standard for the realisation of air kerma for high energies consisted of three graphite-walled cylindrical cavity ionisation chambers. These chambers have become increasingly unreliable in recent years and a project was started at the National Physical Laboratory to replace them. Spherical graphite-walled cavity ionisation chambers PS5-1 and PS5-2 have been constructed and commissioned and this report summarises the factors for the new chambers.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | IR 41 |
| Subjects: | Ionising Radiation > Dosimetry |
| Divisions: | Chemical, Medical & Environmental Science |
| Last Modified: | 27 Feb 2018 15:09 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/7732 |
![]() |
Tools
Tools