Report on dose measurements on the MRI-linac at Liverpool Hospital (Australia) performed by NPL

Billas, I; Homer, M; Duane, S (2018) Report on dose measurements on the MRI-linac at Liverpool Hospital (Australia) performed by NPL. NPL Report. IR 46

[thumbnail of IR46.pdf]
Preview
Text
IR46.pdf - Published Version

Download (1MB) | Preview

Abstract

This report describes measurements of the radiation dose from the MRI-linac at Liverpool Hospital (Australia) using a selection of detectors (such as ionisation chambers, alanine and Gafchromic EBT-3 films) in order to calibrate the machine output. Absorbed dose to water calibration coefficients for the ionisation chambers used in the constant magnetic field (1 T) of the MRI-Linac machine were also measured. Together with similar measurements performed at NPL in a 0 T magnetic field over a range of beam qualities, the corrections required to measure the absorbed dose in the presence of a magnetic field using these detectors was determined.