Dean, J C J; Phillips, H C; Burgess, P*; Doyle, P*; Simpson, J* (2015) The examination, testing and calibration of tritium-in-air monitors for radiation protection. Measurement Good Practice Guide. 114
|
Text (Measurement Good Practice Guide No. 114)
MGPG114.pdf - Published Version Download (1MB) |
Abstract
This Good Practice Guide has been written by members of the Airborne Radioactivity Monitoring Users’ Group in collaboration with the radiation user community. The guide recommends procedures to examine, test and calibrate tritium-in-air monitoring equipment. The guide provide guidelines for Type Tests, Tests Before First Use, Periodic Tests, as well as Function check, Re-tests after repair/adjustment and Pre-calibration tests of the instruments. It also discusses Specific tests, including tests on the radiation detection assembly for portable tritium-in-air monitors. The guide does not cover the use of tritium bubblers. The test procedures recommended in this guide are not legally binding, but are based on current accepted good practice.
| Item Type: | Report/Guide (Measurement Good Practice Guide) |
|---|---|
| Notes: | NPL made every effort to ensure all information contained in this Good Practice Guide was correct at time of publication. NPL is not responsible for any errors, omissions or obsolescence, and does not accept any liability arising from the use of these Good Practice Guides. |
| Subjects: | Ionising Radiation Ionising Radiation > Radioactivity |
| Last Modified: | 20 Feb 2026 14:05 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/6646 |
![]() |
Tools