Roberts, N J; Horwood, N A; McKay, C J (2014) Photon doses in NPL standard neutron fields. Radiat. Prot. Dosim., 161 (1-4). pp. 157-160.
Full text not available from this repository.Abstract
Standard neutron fields are invariably accompanied by a photon component due to the neutron generating reactions and secondary neutron interactions in the surrounding environment. A set of energy-compensated GM tubes and EPD electronic personal dosemeters have been used to measure the photon dose rates in a number of standard radionuclide and accelerator-based neutron fields. The GM tubes were first characterised in standard radioisotope and x-ray photon fields, and then modelled using MCNP to determine their photon dose response as a function of energy. Values for the photon to neutron dose equivalent ratios are presented and compared with other published values.
| Item Type: | Article |
|---|---|
| Keywords: | GM tubes, Electronic Personal Dosemeters, Radionuclide neutron sources, Photon dose, Monoenergetic neutrons |
| Subjects: | Ionising Radiation Ionising Radiation > Neutron Metrology |
| Identification number/DOI: | 10.1093/rpd/nct249 |
| Last Modified: | 02 Feb 2018 13:14 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/6373 |
![]() |