Hawkes, N P; Gamage, K A A*; Taylor, G C (2010) Digital approaches to field neutron spectrometry. NPL Report. IR 23
|
Text
IR23.pdf Download (2MB) |
Abstract
Miniature digitizing electronics have been coupled to a detector based on an organic liquid scintillator to form a digital neutron and gamma spectrometer. The digitizing electronics are physically small enough to be packaged with a detector to form an easily portable unit. The spectrometer was exposed to a selection of the well-characterised radiation fields available at NPL, and the performance of three different neutron / gamma discrimination algorithms were compared. The energy resolution of the system was also assessed. Conclusions are presented on the best algorithm to use in such a system, and on the performance and ease of use of the spectrometer in comparison with conventional analogue systems.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | IR 23 |
| Keywords: | Digital processing, neutron spectrometry, field spectrometry, liquid scintillator, digital pulse shape discrimination |
| Subjects: | Ionising Radiation Ionising Radiation > Neutron Metrology |
| Last Modified: | 02 Feb 2018 13:14 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/4758 |
![]() |