McKay, C J; Burgess, P; Hawkes, N P; Kelly, M; Taylor, G C; Thomas, D J (2009) Photon does in NPL standard radionuclide neutron fields. NPL Report. IR 12
|
Text
IR12.pdf Download (641kB) |
Abstract
The photon dose rates from a range of radionuclide neutron sources have been measured using halogen-quenched Geiger-Müller detectors and EPD MkII and EPD-N2 electronic personal dosemeters. The photon spectrum from each source was calculated from available literature, corrected for self-attenuation within the sources, and used to calculate the Geiger-Müller detector response for each source. EPD and EPD-N2 photon dose measurements on phantom agreed well with the Geiger-Müller detector measurements.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | IR 12 |
| Keywords: | Radionuclide neutron sources, Mixed radiation fields |
| Subjects: | Ionising Radiation Ionising Radiation > Neutron Metrology |
| Last Modified: | 02 Feb 2018 13:15 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/4308 |
![]() |