Williams, A J; Burns, D T; McEwen, M R (1993) Measurement of the specific heat capacity of the electron-beam grahite calorimeter. NPL Report. RSA(EXT) 40
|
Text
RSA40.pdf Download (6MB) |
Abstract
This report describes the design and construction of apparatus to measure the specific heat capacity of solid materials in the temperature range IB-32°C. The apparatus is designed to minimize heat losses from the sample in order to measure the specific heat to an accuracy of better than :to.2% at the 95% confidence level. The results of a set of measurements on the graphite core of the electron-beam primary-standard radiation calorimeter are presented, along with a measurement of the effect of a large accumulated radiation dose on the specific heat of graphite.