Giblin, S P; Lorusso, G (2026) Precision measurement of ionization current from a long half-life radioactive source. Review of Scientific Instruments, 97 (6). 063302 ISSN 0034-6748
Full text not available from this repository.Abstract
We present precision measurements of the ionization current produced by a sample of carbon-14. Using specialized ammeters with a gain uncertainty of less than a part per million (ppm), the decay of the ∼400 pA ionization current, ∼1 fA/week, is resolved by the measurement system for experimental runs of 50 days. When the apparatus is temperature controlled to ±0.05 °C, fluctuations in instrumental offsets are suppressed, and the random uncertainty is limited by Poisson noise, reaching ∼2.5 ppm of the ionization current for 24 h of data integration. We discuss the possibility of a measurement of the carbon-14 half-life by directly observing the decay in the ionization current.
| Item Type: | Article |
|---|---|
| Subjects: | Ionising Radiation > Radioactivity |
| Divisions: | Quantum Technologies |
| Identification number/DOI: | 10.1063/5.0328925 |
| Last Modified: | 15 Oct 2026 14:47 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/10565 |
![]() |