First microdosimetric measurements with a tissue-equivalent proportional counter at the MedAustron ion-beam therapy facility

Barna, S; Meouchi, C; Magrin, G; Bianchi, A; Conte, V; Selva, A; Stock, M; Resch, A F; Georg, D; Palmans, H (2023) First microdosimetric measurements with a tissue-equivalent proportional counter at the MedAustron ion-beam therapy facility. Radiation Protection Dosimetry, 199 (15-16). pp. 1973-1978. ISSN 0144-8420

Full text not available from this repository.

Abstract

The aim of this work is to present the first microdosimetric spectra measured with a miniaturised tissue-equivalent proportional counter in the clinical environment of the MedAustron ion-beam therapy facility. These spectra were gathered with a 62.4-MeV proton beam and have been compared with microdosimetric spectra measured in the 62-MeV clinical proton beam of the CATANA beam line. Monte Carlo simulations were performed using the Geant4 toolkit GATE and a fully commissioned clinical beam line model. Finally, similarities and discrepancies of the measured data to simulations based on a simple and complex detector geometry are discussed.

Item Type: Article
Keywords: microdosimetry, proton beam, Monte Carlo, mini tissue-equivalent proportional counter
Subjects: Ionising Radiation > Dosimetry
Divisions: Medical, Marine & Nuclear
Identification number/DOI: 10.1093/rpd/ncac252
Last Modified: 11 May 2026 13:45