Barna, S; Meouchi, C; Magrin, G; Bianchi, A; Conte, V; Selva, A; Stock, M; Resch, A F; Georg, D; Palmans, H (2023) First microdosimetric measurements with a tissue-equivalent proportional counter at the MedAustron ion-beam therapy facility. Radiation Protection Dosimetry, 199 (15-16). pp. 1973-1978. ISSN 0144-8420
Full text not available from this repository.Abstract
The aim of this work is to present the first microdosimetric spectra measured with a miniaturised tissue-equivalent proportional counter in the clinical environment of the MedAustron ion-beam therapy facility. These spectra were gathered with a 62.4-MeV proton beam and have been compared with microdosimetric spectra measured in the 62-MeV clinical proton beam of the CATANA beam line. Monte Carlo simulations were performed using the Geant4 toolkit GATE and a fully commissioned clinical beam line model. Finally, similarities and discrepancies of the measured data to simulations based on a simple and complex detector geometry are discussed.
| Item Type: | Article |
|---|---|
| Keywords: | microdosimetry, proton beam, Monte Carlo, mini tissue-equivalent proportional counter |
| Subjects: | Ionising Radiation > Dosimetry |
| Divisions: | Medical, Marine & Nuclear |
| Identification number/DOI: | 10.1093/rpd/ncac252 |
| Last Modified: | 11 May 2026 13:45 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/10388 |
![]() |