Assessment of the impact on the PADC neutron dosimetry system of the new operational dose quantities proposed in ICRU Report 95

Moreno, B; Million, M; Thomas, D; Taylor, G (2023) Assessment of the impact on the PADC neutron dosimetry system of the new operational dose quantities proposed in ICRU Report 95. Radiation Protection Dosimetry, 199 (15-16). pp. 1725-1728.

Full text not available from this repository.

Abstract

The International Commission on Radiation Units and Measurements, ICRU, have recommended, in ICRU Report 95, the adoption of new operational quantities for use in radiation protection monitoring. Their implementation will pose problems for instrument manufacturers supplying area dosemeters, and for individual monitoring services providing personal dosemeters for radiation workers. The problems arise for photon, neutron, and beta dosimetry, but this paper deals solely with personal neutron dosemeters based on PADC plastic. It presents information on the problems to be expected based on data from calibrations of actual personal neutron dosemeters in terms of the existing dose quantities.

Item Type: Article
Subjects: Ionising Radiation > Neutron Metrology
Divisions: Medical, Marine & Nuclear
Identification number/DOI: 10.1093/rpd/ncac212
Last Modified: 16 Feb 2024 14:51
URI: https://eprintspublications.npl.co.uk/id/eprint/9932