< back to main site

Publications

The NPL air kerma primary standard PS5-1/PS5-2 for 137Cs and 60CO: summary of factors

Bass, G A; Duane, S; Galer, S E; Homer, M J; Kelly, M; Manning, J W; Maughan, D J; Nutbrown, R F; Sander, T; Shipley, D R; Snaith, J A D (2017) The NPL air kerma primary standard PS5-1/PS5-2 for 137Cs and 60CO: summary of factors. NPL Report. IR 41

[thumbnail of IR41.pdf]
Preview
Text
IR41.pdf - Published Version

Download (325kB) | Preview

Abstract

Since the 1950s the UK primary standard for the realisation of air kerma for high energies consisted of three graphite-walled cylindrical cavity ionisation chambers. These chambers have become increasingly unreliable in recent years and a project was started at the National Physical Laboratory to replace them. Spherical graphite-walled cavity ionisation chambers PS5-1 and PS5-2 have been constructed and commissioned and this report summarises the factors for the new chambers.

Item Type: Report/Guide (NPL Report)
NPL Report No.: IR 41
Subjects: Ionising Radiation > Dosimetry
Divisions: Chemical, Medical & Environmental Science
Last Modified: 27 Feb 2018 15:09
URI: https://eprintspublications.npl.co.uk/id/eprint/7732
View Item