Billas, I; Homer, M; Duane, S (2018) Report on dose measurements on the MRI-linac at Liverpool Hospital (Australia) performed by NPL. NPL Report. IR 46
Preview |
Text
IR46.pdf - Published Version Download (1MB) | Preview |
Abstract
This report describes measurements of the radiation dose from the MRI-linac at Liverpool Hospital (Australia) using a selection of detectors (such as ionisation chambers, alanine and Gafchromic EBT-3 films) in order to calibrate the machine output. Absorbed dose to water calibration coefficients for the ionisation chambers used in the constant magnetic field (1 T) of the MRI-Linac machine were also measured. Together with similar measurements performed at NPL in a 0 T magnetic field over a range of beam qualities, the corrections required to measure the absorbed dose in the presence of a magnetic field using these detectors was determined.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | IR 46 |
| Subjects: | Ionising Radiation > Dosimetry |
| Divisions: | Chemical, Medical & Environmental Science |
| Last Modified: | 27 Feb 2018 14:59 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/7731 |
![]() |