Martin, N A; Goody, B A; Wang, J; Milton, M J T (2012) Accurate and adjustable calibration gas flow by switching permeation and diffusion devices. Meas. Sci. Technol., 23 (10). 105005
Full text not available from this repository.Abstract
A new method for generating adjustable concentrations of calibration gases from permeation and diffusion devices is described. The method achieves high accuracy by maintaining the source device at a constant temperature and pressure and in a constant gas flow. It involves switching the gas stream and subsequently smoothing it to generate dilutions of up to 100 with a relative uncertainty of ± 2%. Its dynamic performance has been modelled and its performance under steady-state conditions has been validated.uncertainty of ± 2%. Its dynamic performance has been modelled and its performance under steady-state conditions has been validated.