Milton, M J T; Wang, J; Harris, P M (2002) Implementation of isotop dilution mass spectrometry with one, two and three reverse steps. NPL Report. COAM 7
|
Text
COAM7.pdf Download (1MB) |
Abstract
An extension to the established methods for analysing an unknown sample by isotope dilution mass spectrometry (IDMS) is proposed. The new method simplifies the presentation of established methods by introducing the 'isotope dilution curve'. This shows how a method can be developed that does not require the measurement of the isotope ratio of the highly enriched spike and therefore limits the range of isotope ratios that have to be measured accurately.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | COAM 7 |
| Keywords: | isotope dilution mass spectrometry (IDMS), two- step IDMS, isotope ratio mass spectrometry, carbon dioxide |
| Subjects: | Analytical Science |
| Last Modified: | 02 Feb 2018 13:17 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/2202 |
![]() |