Perkin, M (2000) Report on a comparison of pressure standards in the range 10 MPa to 100 MPa. NPL Report. CMAM 54
![]() |
Text
cmam54.pdf Download (191kB) |
Abstract
A comparison in the hydraulic pressure range between 10 MPa and 100 MPa was undertaken as EUROMET Project 389. The comparison was originally between five national metrology laboratories but was subsequently expanded to include a further ten participants. The pilot laboratory was the National Physical Laboratory. This report presents the results obtained by each laboratory for the effective area of the piston-cylinder assembly used in the comparison. The results of all participants were found to be consistent with each other and with the calculated reference value, within their claimed uncertainties, at all pressures. The results are therefore considered to be very satisfactory.
Item Type: | Report/Guide (NPL Report) |
---|---|
NPL Report No.: | CMAM 54 |
Subjects: | Engineering Measurements Engineering Measurements > Mass, Force, Pressure |
Last Modified: | 02 Feb 2018 13:17 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/1584 |
Actions (login required)
![]() |