Williams, A J; McEwen, M R; Burns, D T (1998) A new beam current monitor for the NPL linear accelerator facilty. NPL Report. CIRM 12
|
Text
CIRM12.pdf Download (3MB) |
Abstract
A new beam current monitor has been designed and built for the NPL linear accelerator. The output response of the new beam current monitor is relatively free from RF interference: it has a fast rise time of -10 ns, and is stable enough to be calibrated for absolute charge measurements. The uncertainty on these charge calibrations is estimated to be ±0.3% (1s).
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | CIRM 12 |
| Subjects: | Ionising Radiation Ionising Radiation > Dosimetry |
| Last Modified: | 10 Sep 2018 14:21 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/983 |
![]() |
Tools