Heaton, J A (1994) Specification for a secondary standard dosemeter for diagnostic or mammographic radiology. NPL Report. RSA(EXT) 52
|
Text
rsa52.pdf Download (3MB) |
Abstract
A performance specification for dosemeters intended for use as secondary standards for the measurement of X-ray air kerma for diagnostic or mammographic beam qualities is proposed. Important influence quantities are listed and the performance requirements given for ionisation chamber, measuring assembly and stability check device. Consideration is also given to the "useability" of the secondary standard and the security of any external computer software used operationally with the instrument.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | RSA(EXT) 52 |
| Subjects: | Ionising Radiation Ionising Radiation > Dosimetry |
| Last Modified: | 02 Feb 2018 13:19 |