< back to main site

Publications

The NPL high temperature radiation thermometry calibration facility.

Machin, G (1997) The NPL high temperature radiation thermometry calibration facility. NPL Report. CBTM 3

[img] Text
cbtm3.pdf

Download (6MB)

Abstract

The NPL hig temperature radiation thermometry calibration facility consisting of high tempera re blackbody sources and a standard transfer pyrometer is described. The perform ce of the system and the uncertainty budget for the scale realisation is given. It is sh wn that using a suitably calibrated transfer pyrometer, it is possible to realise and. ercompare the ITS-90 above the silver point with a 2s uncertainty of better than O. 5% of temperature.

Item Type: Report/Guide (NPL Report)
NPL Report No.: CBTM 3
Subjects: Engineering Measurements
Engineering Measurements > Thermal
Last Modified: 02 Feb 2018 13:18
URI: http://eprintspublications.npl.co.uk/id/eprint/812

Actions (login required)