Billas, Ilias; Duane, Simon (2018) Report on dose measurements on the Elekta Unity MR linac at the Royal Marsden Hospital performed by NPL. NPL Report. IR 49
Preview |
Text (NPL Report IR 49)
IR49.pdf - Published Version Download (475kB) | Preview |
Abstract
This report describes measurements of the radiation dose from the MR-linac at The Royal Marsden Hospital, Sutton using a selection of detectors (such as ionisation chambers and alanine) in order to calibrate the machine output. Absorbed dose to water calibration coefficients for the ionisation chambers used in the MR-linac machine when the constant magnetic field was either on (1.5 T) or off (0 T) were also determined. Together with similar measurements performed at NPL in a 0 T magnetic field over a range of beam qualities, the corrections required to measure the absorbed dose in the presence of a magnetic field using these detectors can be determined.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | IR 49 |
| Subjects: | Ionising Radiation > Dosimetry |
| Divisions: | Chemical, Medical & Environmental Science |
| Last Modified: | 29 Oct 2018 10:25 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/8117 |
![]() |