Sutton, G J M (2016) The NPL standard flame: operating instructions. NPL Report. ENG 65
|
Text
ENG65.pdf Download (8MB) |
Abstract
This report describes the operating instructions for the NPL portable standard flame system based on the Hencken flat flame burner using dry laboratory air and industrial grade (95% nominal purity) propane. It also includes a summary of the temperature field above the burner for equivalence ratios from phi = 0.8 to phi = 1.4, at an airflow rate of 23 SLPM.
| Item Type: | Report/Guide (NPL Report) |
|---|---|
| NPL Report No.: | ENG 65 |
| Keywords: | standard, flame, combustion, temperature, propane, species |
| Subjects: | Engineering Measurements Engineering Measurements > Thermal |