Acoustic gas thermometry: models and corrections for argon from 90 K to 1500 K.

Sutton, G J M (2014) Acoustic gas thermometry: models and corrections for argon from 90 K to 1500 K. NPL Report. ENG 49

[img] Text
ENG49.pdf

Download (1MB)

Abstract

This NPL report describes the models and corrections required for high precision acoustic gas thermometry using argon gas as the thermometric fluid over the temperature range 90 K < T < 1500 K. Where corrections require a specific geometry and or resonator material, it is assumed that the measurements are performed using a quasi-spherical acoustic resonator made of copper tuned to the radial acoustic modes.

Item Type: Report/Guide (NPL Report)
NPL Report No.: ENG 49
Keywords: acoustic, thermometry, absolute, temperature, boltzmann, thermodynamic, corrections
Subjects: Engineering Measurements
Engineering Measurements > Thermal