Abstract

Two Pt/Pd thermocouples have been manufactured industrially at CCPI-Europe to a robust design developed by NPL. The first has been exposed to temperatures up to 1000 °C for 93 operating hours and the second exposed to higher temperatures (up to 1300 °C) for 327 operating hours, over the course of a year. No significant drift is observed in the temperature measurement for either thermocouple, when compared against reference Type R thermocouples. This work demonstrates the long-term stability and reliability of Pt/Pd thermocouples in an industrial calibration environment with this new robust design.

Item Type: Article
Keywords: Pt/Pd, thermocouple, temperature, industrial
Subjects: Engineering Measurements
Engineering Measurements > Thermal
Identification number/DOI: 10.1063/1.4819600
Last Modified: 02 Feb 2018 13:14