Elliott, C J; Pearce, J V; Machin, G; Ford, T*; Hicks, K* (2013) Pt/Pd thermocouple resilience over 327 operating hours in an industrial calibration laboratory. AIP Conf. Proc., 1552. pp. 549-553.
Full text not available from this repository.Abstract
Two Pt/Pd thermocouples have been manufactured industrially at CCPI-Europe to a robust design developed by NPL. The first has been exposed to temperatures up to 1000 °C for 93 operating hours and the second exposed to higher temperatures (up to 1300 °C) for 327 operating hours, over the course of a year. No significant drift is observed in the temperature measurement for either thermocouple, when compared against reference Type R thermocouples. This work demonstrates the long-term stability and reliability of Pt/Pd thermocouples in an industrial calibration environment with this new robust design.
| Item Type: | Article |
|---|---|
| Keywords: | Pt/Pd, thermocouple, temperature, industrial |
| Subjects: | Engineering Measurements Engineering Measurements > Thermal |
| Identification number/DOI: | 10.1063/1.4819600 |
| Last Modified: | 02 Feb 2018 13:14 |
| URI: | https://eprintspublications.npl.co.uk/id/eprint/5952 |
![]() |