Johansson, L C; Sephton, J P (2010) Validation of a new TDCR system at NPL. Appl. Radiat. Isot., 68 (7-8). pp. 1537-1539.
Full text not available from this repository.Abstract
A new Triple to Double Coincidence Ratio (TDCR) system has been established at NPL. The system incorporates a spherical optical chamber, a manual sample changing facility and an integral light-tight housing. A 6 inch diameter NaI(Tl) detector has been incorporated to allow 4pb-g coincidence measurements to be performed in parallel. Details of the detectors, the electronics and the overall TDCR system are given. Validation measurements using suitable low-energy beta and electron capture nuclides, i.e. H-3 and Fe-55 have been performed. The highest efficiency achieved with the H-3 solution in Ultima Gold AB in a glass vial was 53%. This indicates that the optical chamber is performing well. After varying the efficiency by de-focusing the PM tubes, the activity of the sources was determined. The activity concentration determined with TDCR agreed with certified values within the range of uncertainties. Further results from validation measurements and corresponding uncertainty budgets are presented.
Item Type: | Article |
---|---|
Subjects: | Ionising Radiation Ionising Radiation > Radioactivity |
Last Modified: | 02 Feb 2018 13:15 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/4703 |
Actions (login required)
![]() |
View Item |