< back to main site

Publications

High accuracy dual lens transmittance measurements.

Cheung, J Y; Gardner, J*; Migdall, A*; Polyakov, S*; Ware, M* (2007) High accuracy dual lens transmittance measurements. Appl. Opt., 46 (22). pp. 5396-5403.

Full text not available from this repository.

Abstract

We show how to determine the transmittance of short focal length lenses (f = 19 m and f = 25 mm, in this case) with a combined uncertainty of 3 parts in 10^4 or better by measuring transmittances of lens pairs of a set of three or more lenses with the same nominal focal length. Uncertainties are minimized by optimizing the radiometric design of the setup and the measurement procedue. The technique is particularly useful in systems where the detector acceptance angle limits the beam geometry to relatively collimated beams.

Item Type: Article
Keywords: radiometry, transmittance
Subjects: Optical Radiation and Photonics
Optical Radiation and Photonics > Materials and Appearance
Last Modified: 02 Feb 2018 13:15
URI: http://eprintspublications.npl.co.uk/id/eprint/4037