Keightley, J D; Watt, G C* (2002) Digital Coincidence Counting (DCC) and its use in the corrections for out-of-channel gamma events in 4pß- coincidence counting. Appl. Radiat. Isot., 56 (1-2). pp. 205-210.
Full text not available from this repository.Abstract
The Digital Coincidence COunting (DCC) system developed by NPL and ANSTO is breifly described along with its benefits in the data collection and processing for 4pß- coincidence counting technique of radionuclide standardization. Onr of these benefits is the automatic detection of and correction for out-of-channel coincidences in the Computer Discrimination method. Where the criteria for the use of the Cox-Isham/Smith correction formuae for deadtimes and resolving times are not met, a generalized approximation based on the work of Campion is suggested.
Item Type: | Article |
---|---|
Keywords: | Digital Coincidence Counting (DCC), acquistion hardware, software analysis, 4pß- coincidence counting |
Subjects: | Ionising Radiation Ionising Radiation > Radioactivity |
Last Modified: | 02 Feb 2018 13:17 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/2071 |