Uncertainty evaluation of the relative humidity facility at NPL.

Stevens, M (1998) Uncertainty evaluation of the relative humidity facility at NPL. NPL Report. CMAM 15

[img] Text
cmam15.pdf

Download (6MB)

Abstract

A relative humidity calibration facility has been constructed and validated at the National Physical Laboratory. This facility uses the present NPL humidity generators to provide gas of a known dew point to one of two temperature-controlled chambers. In this way a known relative humidity is generated in the chamber. The instrument to be calibrated is placed in the temperature-controlled chamber. The range of temperatures available at present is from 1 °C to 80 °C with a range of relative humidity of 2 %rh to 98 %rh over most of the temperature range. The uncertainty with which this new facility will operate is in general betiter than 0.5 %rh. This project has been carried out as part of the DTI's National Measurement System Programme for Mass, 1996 to 1999, project 4, Humidity Standards.

Item Type: Report/Guide (NPL Report)
NPL Report No.: CMAM 15
Subjects: Engineering Measurements
Engineering Measurements > Thermal
Last Modified: 10 Sep 2018 14:06
URI: http://eprintspublications.npl.co.uk/id/eprint/1000