Skinner, J (2022) Benchmarking electronic calibration of USB-enabled network analysers. In: ARMMS RF & Microwave Society Conference 2022, 25-26 April 2022, Thame, Oxfordshire, UK.
Full text not available from this repository.Abstract
These days, many instrument calibrations are performed automatically – for example, using electronic calibration units. This includes calibration of vector network analysers (VNAs) where the calibration standards are generated electronically rather than using mechanical devices. However, verifying performance – particularly accuracy – of these electronic units is more difficult due to the “black box” nature of these devices. This paper describes a benchmarking exercise of a new-to-market electronic calibration unit, designed specifically for use with a portable USB-driven VNA. This includes a comparison of performance with mechanical calibration kits, and a comparison with the most accurate measurements available, obtained using primary national standards.
Item Type: | Conference or Workshop Item (Paper) |
---|---|
Keywords: | VNA ; Electronic calibration; e-cal ; pico ; usb-enabled |
Subjects: | Electromagnetics > RF and Microwave |
Divisions: | Electromagnetic & Electrochemical Technologies |
Last Modified: | 14 Nov 2023 15:03 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/9861 |