A practical approach to report output factors for the NPL beam data acquisition service. Emphasis in small fields

Silvestre, I; Subiel, A; Dimitriadis, A; Suane, S; Jnaith, J; Homer, M (2016) A practical approach to report output factors for the NPL beam data acquisition service. Emphasis in small fields. NPL Report. IR40

[img] Text
IR40.pdf

Download (984kB)

Abstract

As part of Pinnacle treatment planning commissioning, a practical approach to report output factors for small fields (0.5x0.5 up to 3x3cm2) using the daysi chain was investigated. Experimental field size and detector specific correction factors were also determined. Comparison between the two methods was reported. Output factors for all the relevant field sizes (0.5x0.5 up to 40x40cm2) were determined for the NPL Linac with the Agility head. The results allows for a set of reference data that are also useful for comparison with measurements performed during the NPL beam data acquisition service RD18, for similar linacs and conditions as the ones in the report.

Item Type: Report/Guide (NPL Report)
NPL Report No.: IR40
Keywords: small fields, daisy chain, correction factors
Subjects: Ionising Radiation
Ionising Radiation > Dosimetry
Last Modified: 02 Feb 2018 13:13
URI: http://eprintspublications.npl.co.uk/id/eprint/7538