< back to main site

Publications

A wideband tissue-equivalent liquid for multiband specific absorption rate measurements (850 MHz to 2.5 GHz).

Mouthaan, R; Loader, B G (2016) A wideband tissue-equivalent liquid for multiband specific absorption rate measurements (850 MHz to 2.5 GHz). Electron. Lett., 52 (3). pp. 176-177.

Full text not available from this repository.

Abstract

A tissue-equivalent liquid has been developed that yields permittivity and conductivity values within +/-10% of the IEEE 1528, IEC62209 and FCC OET65-C target 'head' values betweem 850 MHz and 2.5 GHz. The formulation of the new tissue-equivalent liquid is presented alongside measurements of the dielectric properties, thermal properties and density.

Item Type: Article
Keywords: SAR, tissue-equivalent, liquid, broadband
Subjects: Electromagnetics
Electromagnetics > RF and Microwave
Identification number/DOI: 10.1049/el.2015.3707
Last Modified: 02 Feb 2018 13:13
URI: http://eprintspublications.npl.co.uk/id/eprint/6960