Ryding, S*; Cernik, R*; Wooldridge, J; Burnett, T L; Stewart, M; Vecchini, C; Cain, M G; Lennie, A*; Yuan, F J*; Tang, C*; Thompson, P* (2013) Simultaneous measurement ox X-ray powder diffraction and ferroelectric polarisation data as a function of applied electric field at a range of frequencies. Powder Diffr., 28. S220-S227
Full text not available from this repository.Abstract
A variable frequency ferroelectric polarisation measurement system has been designed and built at the UK's Diamond Light Source. The electric field induced phase transitions in Pb(Zr1-xTix)O3 (PZT) have been used to test the facility via in situ measurements of electric polarisation and XRD. Stroboscopic and real time data collection methods on polycrystalline samples were employed as a function of frequency to determine the dynamic ferroelectric response. The system has been shown to deliver XRD patterns of good statistical quality measured over 40 points of a ferroelectric PE loop. The system is now available on station I11 as a user facility at the Diamond Light Source.
Item Type: | Article |
---|---|
Keywords: | in-situ electric field, high resolution powder diffraction |
Subjects: | Advanced Materials Advanced Materials > Functional Materials |
Identification number/DOI: | 10.1017/S0885715613001024 |
Last Modified: | 02 Feb 2018 13:14 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/6022 |
Actions (login required)
View Item |