Humphreys, D A; Miall, J (2013) Traceable measurement of source and receiver EVM using a real-time oscilloscope. IEEE Trans. Instrum. Meas., 62 (6). pp. 1413-1416.
Full text not available from this repository.Abstract
A method to traceably calibrate Wideband Code Division Multiple Access (WCDMA) Error Vector Magnitude (EVM) contributions of a source or receiver is described using a Real-Time Digital Oscilloscope as the reference receiver. The results and uncertainties are presented for a receiver calibrated using four WCDMA sources from different manufacturers. The weighted mean EVM of the receiver at 900 MHz is 0.35 ±0.03 % at 95% confidence. Measurement results at 1780 MHz and 2510 MHz are also reported.
Item Type: | Article |
---|---|
Keywords: | Code division multiaccess, Communications equipment testing, Measurement errors, Oscilloscopes, Uncertainty, Test equipment |
Subjects: | Electromagnetics Electromagnetics > Wireless Communications |
Identification number/DOI: | 10.1109/TIM.2013.2238458 |
Last Modified: | 02 Feb 2018 13:14 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/5814 |