Woods, M J; Ciocanel, M; Keightley, J D (1997) Intercomparison of 123I solution sources in UK Hospitals, 1996. NPL Report. CIRA(EXT)017
Text
cira17.pdf Download (15MB) |
Abstract
During recent years, concerns have been raised within the nuclear medicine field about the accuracy of calibrations for 1231 using radionuclide calibrators. Significant variations in response between different containers have been observed due to the low energy of the associated photon emissions.
To investigate the level of measurement performance for this particular radionuclide, an intercomparison exercise was conducted between the National Physical Laboratory, Amersham International plc and the UK hospital physics community.
The conduct of that intercomparison is described here, additional experimental results are
presented, the results are discussed and recommendations made.
Item Type: | Report/Guide (NPL Report) |
---|---|
NPL Report No.: | CIRA(EXT)017 |
Subjects: | Advanced Materials |
Last Modified: | 02 Feb 2018 13:18 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/570 |
Actions (login required)
View Item |