Humphreys, D A; Akmal, A* (2012) Channel timebase errors for digital sampling oscilloscopes. pp. 520-521.
Full text not available from this repository.Abstract
This paper describes measurement of inter-channel timing errors for two Digital Sampling Oscilloscope designs. Timebase correction algorithms compensate the main error component ±15 ps but a channel specific residual term of up to ±4 ps remains.
Item Type: | Article |
---|---|
Keywords: | Sampling oscilloscopes |
Subjects: | Electromagnetics Electromagnetics > Wireless Communications |
Identification number/DOI: | 10.1109/CPEM.2012.6251032 |
Last Modified: | 02 Feb 2018 13:14 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/5567 |