Naftaly, M; Dudley, R A (2009) Linearity calibration of amplitude and power measurements in terahertz systems and detectors. Opt. Lett., 34 (5). pp. 674-676.
Full text not available from this repository.Abstract
We describe a device consisting of a set of silicon plates to be employed for calibrating the linearity of amplitude/power measurements of terahertz systems and detectors. The device is used to test a time-domain spectrometer, a pyroelectric sensor and a Golay cell. The results demonstrate that the linearity of amplitude/power measurement cannot be assumed but must be experimentally verified.