The NPL air kerma primary standard TH100C for high dose rate 192Ir brachytherapy sources.

Sander, T; Nutbrown, R F (2006) The NPL air kerma primary standard TH100C for high dose rate 192Ir brachytherapy sources. NPL Report. DQL-RD 004

[img] Text
DQL_RD4.pdf

Download (2MB)

Abstract

In 2002 a project was started at the National Physical Laboratory with the aim to set up a new calibration service for dosemeters used to determine the reference air kerma rate of high dose rate 192Ir brachytherapy sources. This report details the construction and the commissioning of the spherical graphite-walled cavity ionisation chamber and the associated measurement set-up which was established as the UK national primary standard for 192Ir in May 2004.

Item Type: Report/Guide (NPL Report)
NPL Report No.: DQL-RD 004
Subjects: Ionising Radiation
Ionising Radiation > Dosimetry
Last Modified: 02 Feb 2018 13:15
URI: http://eprintspublications.npl.co.uk/id/eprint/3749