Rietveld, G*; de la Court, P*; Parkin, G I; Barker, R M (2005) Evaluation of the self-sdjustment routine of a commercial LCR meter. In: BEMC 2005 - British Electromagnetic Measurements Conference, 14-17 November 2005, NPL, UK.
![]() |
Text
BEMC2005-6.pdf Download (244kB) |
Abstract
In a combined project of NMi and NPL we have evaluated the self-adjustment facility of a commercial LCR meter. The evaluation concerned both the hardware as the embedded software that is controlling the adjustment process. The main finding of the evaluation is that the adjustment procedure has a good potential for being the basis of a traceable calibration of the instrument, but that especially the software needs to be significantly improved upon.
Item Type: | Conference or Workshop Item (UNSPECIFIED) |
---|---|
Subjects: | Electromagnetics Electromagnetics > Electrical Measurement |
Last Modified: | 02 Feb 2018 13:16 |
URI: | http://eprintspublications.npl.co.uk/id/eprint/3313 |
Actions (login required)
![]() |
View Item |