NPL contribution to an international comparison of 24.5 keV neutron fluence measurements.

Thomas, D J (2001) NPL contribution to an international comparison of 24.5 keV neutron fluence measurements. NPL Report. CIRM 46

[img] Text
CIRM46.pdf

Download (186kB)

Abstract

This report contains details of the experimental procedure and the analysis techniques used by NPL during an international comparison of neuron fluence production and measurement in the energy region around 24 keV. The exercise involved the calibration of three spheres from a Bonner sphere set in terms of their fluence response to 27.4 and 24.6 keV neutrons produced using the 45Sc(p,n)45Ti reaction.

Item Type: Report/Guide (NPL Report)
NPL Report No.: CIRM 46
Subjects: Ionising Radiation
Ionising Radiation > Neutron Metrology
Last Modified: 02 Feb 2018 13:17
URI: http://eprintspublications.npl.co.uk/id/eprint/1952

Actions (login required)

View Item View Item