< back to main site

Publications

Protocol for evaluation and calibration of on-site pH measuring equipment.

Champion, R*; Brown, R J C; Ficicioglu, F*; Ferrarra, E*; Fisicaro, P*; Mariassy, M*; Spitzer, P* (2008) Protocol for evaluation and calibration of on-site pH measuring equipment. Technical Report.

Full text not available from this repository.

Abstract

This protocol is intended for the calibration and the evaluation of on-site measuring equipment for the determination of pH in water. It gives guidance to the end-user on how to estimate the measurement uncertainty of the pH value following standard procedures. This protocol can be used for on-line and on-site measuring devices over a pH range from pH = 3 to pH = 10 for measurements in water at ionic strength I below 0.3 mol/kg and in a temperature range of 0°C to 50 °C. (Electrolytic conductivity of the solution at 25 °C below 2000 mS/m).

Item Type: Report/Guide (Technical Report)
Subjects: Analytical Science
Analytical Science > Trace Analysis and Electrochemistry
Last Modified: 23 Jul 2018 13:21
URI: http://eprintspublications.npl.co.uk/id/eprint/4230

Actions (login required)